Technique / Immunology / ELISA assay
Edge Effect in MicroWell ELISA
Edge Effect in MicroWell ELISA
Nalge Nunc International. (broken link fixed on March 1, 2005)
Sometimes with ELISA performed in a MicroWell plate unexpectedly higher(or lower) optical densities (O.D.) are measured in the peripheral wells than in the central wells. This phenomenon is called ¡±edge effect¡±.
The most probable causes of this effect are illumination or temperature differences between the peripheral and the central wells.
Light may cause edge effect if the ... ...
URL: Edge Effect in MicroWell ELISA
Contents: original article
Last update 27-May-2003, Rating of 1 votes.